Larger picture |
Ambient air
is compressed in an air-cooled compressor, that can either be
integrated to the generator, or be as a separate compressor.
Compressed air, free from moisture and oil, is led into the unit, and
flows through the carbon molecular sieve (CMS) beds, where oxygen,
carbon dioxide, moisture and some hydrocarbons are adsorbed, and a
stream of clean, dry, high purity nitrogen gas leaves the adsorber
beds to a storage tank. During the CMS beds regeneration period, when
the beds are depressurized to ambient pressure, nitrogen is taken from
the tank. The technology is called pressure swing adsorbtion (PSA).
The adsorption / desorption cycle is controlled by a microprocessor
controller. Nitrogen flow and purity are factory set prior to shipment
of the system, to meet customers
requirements.
-
Microprocessor
controlled
-
Silent,
continuous operation
-
Operation
method PSA CMS
-
Remote
control and alarm
-
Compact
size, easy to install
-
Mono or
dual bed design
|
|
|
| Technical Specifications |
| Model,
gas purity |
NGLCMS
99,5% |
NGLCMS
99,9% |
|
Basic module width mm |
324 |
|
Basic module depth mm |
482 |
|
Basic module Height mm (including
wheels 48 mm) |
820 |
|
Basic moduleWeight kg |
59 |
| Flow
cc/min |
0 - 40000 |
|
Number of basic modules for system |
1 - 2 |
2 - 4 |
|
Buffer tank size for nitrogen L,
mono bed (dual bed) |
100 - 200 ( 40 - 80 ) |
| Dew
Point N2 |
-15 to -70 depending on flow |
| Inlet
/ Outlet pressure barg |
10 - 7 |
6,5 - 5 |
| Power Rating,
control electronics, W / noise level |
8
W / <55 dB at 1 m |
| Integrated compressor or oxygen
analyzer |
Optional |
| Electrical |
230 V / 50 Hz or 110 V / 60 Hz |
|
Please fill out